Skip to main content

PDF


Title: The evaluation of a Philips ammonium ion selective electrode (220/14/T)
Author: D.A. Neville
Author: M.M.I. Harman
Author: P.D. Whalley
Author: A. Chappell
Document Type: Monograph
Abstract:
This report describes the results of an evaluation of a Philips ammonium ion selective electrode The evaluation was undertaken by the NRA (Thames Region) at the Evaluation and Demonstration Facilities at Fobney Mead, Reading and Lea Marston, Birmingham according to an evaluation protocol jointly devised by WRc and the NRA. Generally the electrode was found to be very easy to operate and maintain. The documentation received was clear and well written with instructions for installation, operation and storage. The maintenance requirements of the electrode were low, being a sealed unit, except under certain field conditions, where the water quality was sufficiently poor to necessitate regular cleaning of the electrode to remove foulant. Laboratory trials to determine sensor accuracy established that the total error (quadrature sum of random and systematic errors).
Publisher: National Rivers Authority
Publication Date: 1993
Publication Place: Bristol
Subject Keywords: ExperimentationAmmoniaElectrodesEvaluation
Extent: 51
Permalink: http://www.environmentdata.org/archive/ealit:2280
Total file downloads: 301

Download PDF    Display PDF in separate tab